4.3
CONNECTING TO MEASURING DEVICE
Read safety section of instructions before operating
this product.
METRAFLEX 3000: Plug the output cable of the current sensor
into the voltage measuring input terminals of the multimeter
(black to GND, red to V) and switch it to V AC.
METRAFLEX 3001/3003: Plug each output cable of the current
sensor to the current measuring input terminals "I" on the
MAVOWATT|50.
METRAFLEX 3003 also the phase labelling L1, L2, L3.
METRAFLEX 6001XBL/6003XBL: Plug each output cable of the
current sensor to the current measuring input terminals on the
MAVOWATT 20/30/40/70 or MAVOSYS 10. With METRAFLEX
3003XBL observe the phase labelling.
4.4
CONNECTING TO THE CIRCUIT TO BE
MEASURED
It is strongly recommended to de-energize the circuits to
be
measured
installation. If it is necessary to make connections on
energized circuits they must be made by qualified and
authorized personnel only observing the required safety
precautions.
Never apply the probe around bare conductors with
voltage levels over 1000V (CAT III) or 600V (CAT IV) to
ground.
Always connect electronics unit to measuring device
before installing the flexible measuring head.
Open the probe coupling and slip the measuring head over the
conductor carrying the current to be measured.
Close the probe coupling such that it visibly and audibly snaps
into place.