Commissioning
3.4 Setting the Parameters
4. Trigger <ENTER> to save selection. The [Anomaly Diams] menu will appear.
5. Trigger <Right Arrow> to select "=" sign.
6. Trigger <Up/Down Arrows> to select number of pipe diameters.
7. Trigger <ENTER> to save selection.
8. To return to the top menu level, continue to trigger the <Left Arrow>.
Table 3- 3
Pipe Configuration Option List Definitions
Fully Developed
1 Elbow
Dble Elbow+
Dble Elbow-
Valve
Expander
Reducer
Norm Entry
Header Inlet
Intrusions
Transducer Selection Example for D1H High Precision Transducers
1. At the [Chan/Path Setup] menu, trigger the<Down Arrow> to select [Pick/install Xdcr].
2. Trigger <Right Arrow> then <Down Arrow> to [Transducer Model].
3. Trigger <Right Arrow> sensor.
4. Trigger <Down Arrow>and scroll to [1011HP-T1.]. Trigger <ENTER> to select.
5. Trigger <Down Arrow> to select [Transducer Size].
26
All manuals and user guides at all-guides.com
Use this menu cell to select the pipe configuration that most accurately represents the
upstream pipe condition.
Use this menu cell to enter the number of pipe diameters between the upstream
configuration and the transducer installation.
Options
Definitions
Fully developed flow, as would be expected for very long
straight pipe runs or installation downstream of a flow
condition.
Single 90 degree Elbow upstream of transducer installation.
Double out-of-plane Elbows upstream of transducer
installation.
Double in-plane Elbows upstream of transducer installation.
Not available at this time.
Pipe expansion upstream of transducer installation.
Pipe reduction upstream of transducer installation.
Not available at this time.
Header or pipe manifold upstream of transducer installation.
Not available at this time.
Operating Instructions (Compact), 12/2008, CQO:QSG010 Revision A
SITRANS FUH1010 Quick Start