Installation
i. Installation requirements
Space requirements
The drawing shows the minimum space required for
safety reasons:
Figure i.1
Each drawing has two sets of dimensions:
1
from the wall to the centre of the holes on
the left and at the top of the drawing
2
from the wall to the outline of the cabinet: on
the right and bottom of the drawing
Floor requirements
The fl oor must be:
-
horizontal; +/- 1° tolerance
-
even; tolerance within 2 mm
-
able to bear the weight of the balancer as stated in
Chapter 2 of the Operator's Manual.
The fl oor on which the balancer will be installed should
not transmit vibrations from other devices or from
outside the building. External vibrations may affect the
accuracy of the unit.
Note: The balancer should be positioned on the fl oor
directly. Do not use spacers to fi ll gaps.
If the above conditions are satisfi ed, the balancer does
not need fi xing to the fl oor.
Power supply requirements
Refer to Chapter 2 of the Operator's Manual for power
mains requirements.
WARNING: ENSURE THAT AN APPROVED WALL
MAINS OUTLET IS AVAILABLE.
WARNING: NEVER GUIDE POWER SUPPLY CABLES
OVER THE FLOOR, UNLESS PROTECTED BY AN
APPROVED COVER.