SPECIFICATIONS

Electrical Requirements:
Rated Current Input:
Sound Level:
Altitude:
*Oxygen Output
Concentration Levels:
Maximum Outlet Pressure:
Label
Status
Symbol
SYSTEM FAILURE
Below 73%
73% to 85%
SYSTEM OKAY
O
2
Over 85%
Alternating Current
Type BF Equipment
Running
NOT Running
Attention-Consider Accompanying Documents
DO NOT Smoke
Class II
230 VAC ± 10% (253 VAC/207 VAC), 50 Hz
2.0 A
44 dB Average
Up to 1828 m(6,000 ft) above sea level without
degradation of concentration levels. NOT
RECOMMENDED FOR USE ABOVE
1828 M (6000 FT).
ALL 5LXA/5LXO2A MODELS
93% minimum at 1 to 3 L/min
91% minimum at 4 L/min
87% minimum at 5 L/min (maximum
recommended flow)
* Stated concentration levels
achieved after initial warm-up period
(approximately 30 minutes).
34.5 ± 3.45 kPa
(5 ± 0.5 psi)
5LXO
INDICATORS:
2
Indicator Lights (LED)
RED Continuous Audible Alarm Sieve-GARD
Compressor Shutdown
9
SPECIFICATIONS
YELLOW
GREEN
TM