Cabinet; Humidifier; Troubleshooting Guide - Invacare PLATINUM 5 IRC5LXO2AW Manual De Instrucciones

Concentradores de oxígeno standard
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 25
MAINTENANCE/TROUBLESHOOTING
B. CABINET
UNPLUG the concentrator when cleaning. To avoid electrical
shock, DO NOT remove cabinet.
1. Clean the cabinet with a mild household cleaner and non-abrasive
cloth or sponge.
C. HUMIDIFIER
NOTE: To clean the oxygen humidifier, follow the instructions provided by
the manufacturer. If none are provided, follow these steps:
1. Clean the humidifier every day.
2. Wash it in soapy water and rinse it with a solution of ten (10)
parts water and one (1) part vinegar.
3. Rinse thoroughly with hot water and refill with distilled water to
the level shown on the humidifier.
DO NOT overfill Humidifier Bottle.
CONCENTRATOR

TROUBLESHOOTING GUIDE

SYMPTOM:
PROBABLE CAUSE:
Alarm:
Main Power Loss:
SHORT BEEPS,
1. Power cord NOT
LONG PAUSE
Concentrator
2. No power at outlet.
NOT operating,
3. Tripped circuit breaker.
power switch ON.
Beep......Beep......
Alarm:
System Failure:
1. Unit overheating due
CONTINUOUS
Concentrator
NOT operating,
Power Switch ON.
Beeeeeeeep....
2. Insufficient power at
3. Internal repairs required.
WARNING
WARNING
plugged In.
to blocked air intake.
outlet.
20
SOLUTION:
1. Insert plug into outlet.
2. Inspect house circuit
breakers or fuses. If
problem recurs, use a
different outlet.
3. Call Service Provider.
1a. Remove and clean
cabinet filters.
1b. Move oxygen concentrator at
least 7.5 cm (3 in) away from
walls, draperies or furniture.
2. Do not use extension
cords. Move to another
electrical outlet or circuit.
3. Call Service Provider.

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Platinum 5 irc5lxawIrc5lxo2awIrc5lxaw

Tabla de contenido