Troubleshooting
The green light blinks, but stops after a while.
The green light blinks.
All or parts of the battery scale blink quickly.
Red light + battery scale blink
If unusual sounds are heard:
The lift cannot be started
SystemRoMedic
I N S T R U C T I O N S F O R U S E
TM
Automatic calibration complete
- Manual calibration is required, see section on manual
calibration.
- Check to ensure that no load was applied to the
handlebar when the lift was started e.g., that nothing was
hanging on the handlebar. In that case, remove it and the
green light will stop blinking.
Something is wrong with the lift's electrical system.
Restart EvaDrive. If there is still a problem, contact
Handicare.
- Restart
- Contact Handicare
- Try to locate the source of the sound.
- Take the lift unit out of service and contact Handicare.
- Check the battery switch.
- Check battery connection.
- Connect charger and charge for at least 2 hours.
- Contact Handicare.
11