Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
Proximity fields from RF
wireless communica‐
tions equipment
IEC 61000-4-3
Note: U
is the a.c. mains voltage prior to application of the test level.
T
Radiated RF
IEC 61000-4-3
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption, and reflection from structures, objects and people.
IEC 60601-1-2:2014
Test Level
70% for 25/30 cycles (at
0°)
0% for 250/300 cycles
30 A/m
IEC 60601-1-2:2014,
Table 9
10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
Max foreseeable use
condition per IEC
60601-1-2:2014
30 A/m
IEC 60601-1-2:2014,
Table 9
10 V/m
80 MHz to 6 GHz
80% AM at 1 kHz
40
Electromagnetic Envi‐
ronment Guidance
For use in a typical
domestic, commercial,
or hospital environment.
For use in a typical
domestic, commercial,
or hospital environment.
Portable and mobile RF
communications equip‐
ment should be used no
closer to any part of the
transmitter than the rec‐
ommended separation
distance of 30 cm (12 in).
Field strengths from
fixed RF transmitters, as
determined by an elec‐
tromagnetic site survey,
should be less than the
compliance level in each
frequency range.
Interference may occur
in the vicinity of equip‐
ment marked with the
following symbol: