PLEASE PERFORM THE FOLLOWING STEPS WITH DOOR OPENED
To set the date and time which are shown on the display, proceed as follows:
1. Turn on the autoclave with the "ON/OFF" main switch, located on the right side of the
control panel and keep pressing the "START/STOP" button. After some seconds the
"TIME/MESS." display shows the writing -1-. This means you accessed the first
programming page and you can now release the "START/STOP" button.
The "TEMPERATURE" display shows the blinking seconds. Press the "14A" button, to
access selection of the minutes (MM display). To change the values displayed on "MM"
press the "13A" button or the "16A" button to increase or respectively decrease them.
Now, press the "14A" button, to enter hours selection (H display). The minutes stop
blinking and the hours start blinking. Set the hours to the desired time data pressing the
"13A" button or the "16A" button, to increase or respectively decrease the current
setting.
Press "14A" to enter setting of the day (DATE display). Press "13A" or "16A" to increase
or respectively decrease the current day value. Press "14A" to enter setting of the
month (DATE display). Press "13A" or "16A" to increase or respectively decrease the
current month value.
Press "14A" to enter setting of the year (DATE display). Press "13A" or "16A" to increase
or respectively decrease the current year value.
2. After setting the new date and time, press the "START/STOP" button, to confirm the
data and enter the next programming page. The "TIME/MESS." display shows the
writing -2-. On this page, it is possible to align the atmospheric pressure. The
"PRESSURE" display shows the blinking message 0. Press the "START/STOP" button,
to confirm the new value. This setting is required to align the atmospheric pressure
which changes based on the local altitude. An acoustic alarm warns you when the new
data are saved.
3. Now, the system automatically enter the next programming page -3-. Here it is possible
to set the type of external units connected to the system (TEMPERATURE display).
Pressing "13A" button or the "16A" button to increase or respectively decrease the
number of connected units: 0=None – 1=Internal Printer – 2=External Printer –
3=Computer – 4=Internal Printer+Computer.
4. Press the "START/STOP" button to confirm the new data and enter the following
programming page –4- ( "TIME/MESS." display.). This page shows the autoclave serial
number. Do not change this setting.
change this setting. Turn off the "ON/OFF" main switch, to exit this menu.
16
SETTING THE DATE AND TIME