Descargar Imprimir esta página

DENTSPLY Cavitron SPS Instrucciones De Uso página 6

Ocultar thumbs Ver también para Cavitron SPS:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 33
. Electro-Magnetic Compatibilty Precautions
Guidance And Manufacturer's Declaration - Electromagnetic Emissions
The Ultrasonic Scaler model G135 is intended for use in the electromagnetic environment specified
below. The customer or the user of the Ultrasonic Scaler should assure that it is used in such
an environment.
Emissions test
RF emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance And Manufacturer's Declaration – Electromagnetic Immunity
The Model G135 is intended for use in the electromagnet environment specified below. The customer or the
end user of the Model should assure that it is used in such an environment.
IEC 60601
Immunity test
Test level
+6 kV contact
Electrostatic
-
+8 kV air
discharge (ESD)
-
IEC 61000-4-2
3 A/m
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
ENGLISH • 
Compliance
Group 1
The Ultrasonic Scaler uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
The Ultrasonic Scaler is suitable for use in all establishments,
Class B
including domestic establishments and those directly
connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Class A
Not Applicable
Not Applicable
The Ultrasonic Scaler is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Compliance level
+6 kV contact
-
+8 kV air
-
3 A/m
Electromagnetic environment – guidance
Electromagnetic environment -
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.

Publicidad

loading