FUSE REPLACEMENT
1. Turn the unit off and remove the power cord from both the wall outlet and
the Quantum 2000 Light Source.
2. Use a small screwdriver to remove the fuse holder below the power cord
connector by releasing the spring-loaded snap connections.
3. Pull out the fuse holder.
4. Remove and replace with new fuses of the same type. There is no
required orientation of the fuses. Be sure to replace both fuses.
5. Reinsert the fuse holder by pressing it into the power connector until it
snaps into place.
6. If the new fuse blows, immediately contact Stryker Technical Service.
F u s e s
Snap
Connections 7
Fuse Holder
8