Electromagnetic Compatibility Information

Guidance and manufacture's declaration – electromagnetic emissions – for all EQUIPMENT and SYSTEMS
Guidance and manufacture's declaration – electromagnetic emission
The BC81 blood pressure monitor is intended for use in the electromagnetic environment specified below. The customer of the user of the
BC81 blood pressure monitor should assure that it is used in such and environment.
Emission test
Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1
RF emission CISPR 11
Class B
Guidance and manufacture's declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS
Guidance and manufacture's declaration – electromagnetic immunity
The BC81 blood pressure monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the
BC81 blood pressure monitor should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Power frequency (50Hz)
magnetic field
IEC 61000-4-8
The BC81 blood pressure monitor uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby electronic equip-
ment.
The BC81 blood pressure monitor is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
IEC 60601
Compliance
test level
level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floor are covered with
synthetic material, the relative humidity should be at least 30%.
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
101