Attaching the Cushion
1. Push the cushion into the frame towards the front and press into place
2. The bottom of the cushion will match the alignment cutouts of mask frame
Attaching the Elbow and Tubing
Push the elbow into the cushion until the Quick Release Tabs click into place
Attaching the Headgear
1. Hold the crown strap upward.
2. Make sure the wide strap marked with "Philips Respironics" is outward.
3. Slide the headgear tabs into the longer frame arm slots.
4. Connect the headgear clips to the frame or slide the tabs into the shorter frame arm slots if
not using the headgear clips.
Optional: Disassembly/Assembly of Magnetic Headgear Clips
If your Wisp mask contains magnetic headgear clips
14
Removing the Magnetic Headgear Clips
1.
Pull the front end of the magnetic clip away from the back circular end of the clip to
disconnect the magnets.
2.
Push the circular end back and out of the mask frame bottom slot.
the mask frame for easier removal.
Attaching the Magnetic Headgear Clips
1.
Push the back circular end of the magnetic headgear clip into the mask frame bottom slot
until it clicks into place.
2.
Connect the front end of the magnetic clip to the back circular end. The clip will snap into
place when the magnetic ends are connected.
Philips Respironics System One Resistance Control
This mask when combined with a Philips Respironics System One device, provides optimal
resistance compensation. This mask value is X1 and should be set by the provider.
Notes:
• Compare the mask to the device. See the device manual if the values do not match.
• System One is not compatible with masks that use a separate exhalation device.
Specifications
Warning: The technical specifications of the mask are provided for the healthcare
professional to determine if it is compatible with the CPAP or bi-level therapy device. If
used outside these specifications, or if used with incompatible devices, the mask may be
uncomfortable, the seal of the mask may not be effective, optimum therapy may not be
achieved, and leak, or variation in the rate of leak, may affect device function. The pressure
flow curve shown below is an approximation of expected performance. Exact measurements
may vary.