Descargar Imprimir esta página

BRAND Transferpette-8 Instrucciones De Manejo página 76

Ocultar thumbs Ver también para Transferpette-8:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 127
10 Maintenance
Complete piston shaft system with central guide axis (Z) connected to the pipetting housing cover [II]
(channel identification can be read here).
B
Piston–shaft units inserted in the frame [I] of the pipetting unit. The units consist of the piston (with
seal) (B') and the stem (with O-ring) (B"), which can be disassembled for cleaning, lubrication, and re-
placement.
C
Pipetting housing connected to the pipetting housing cover [II] by two sliding catches.
Cleaning instructions
a.
Clean individual shafts and pistons (only these parts) with soap solution or isopropanol. Then
rinse with distilled water.
b.
Allow the parts to completely dry and cool. Liquid residues in the shafts result in deviations in ac-
curacy.
c.
Re-grease pistons with a very thin coat of the grease supplied. For the central guide axis (Z), use
only the prescribed fluorine static grease.
d.
After assembling the unit, carry out a reference run (rEF).
Removing shafts and pistons for cleaning or replacement
b.
c.
d.
76
Operating manual
a.
Unscrew pipetting unit: Separate the pipetting unit
from the handle.
b.
Remove the pipetting housing: Pull the sliding catches
of the pipetting housing cover out to the side as far as
they will go and remove the pipetting housing.
c.
Move the piston to the lower position: Push the plunger
down until the pistons are in the lowest position.
d.
Remove the shaft attachment: Pull out the shaft attach-
ment from the side.
Transferpette® -8/-12 electronic 997200

Publicidad

loading