Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfecting machine without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, microbiological, at least of drinking water quality)
*Recommended: BBraun Helimatic Cleaner alcaline
►
Insert the product in its proper position in the
ECCOS fixation.
►
Screw the rinsing adapter (GA672211) on the prod-
uct in the mount 17.
►
Then connect the interior rinse device and connect
to the rinse connector of the cleaning/disinfection
device/rinse cart.
►
Attach the Kirschner wire protection sleeve 15 to a
rinse hose.
►
After automatic cleaning/disinfection, check visible
surfaces for residues and repeat the cleaning/disin-
fection process as needed.
6.8
Inspection, maintenance and
checks
►
Allow the product to cool down to room tempera-
ture.
►
Inspect the product after each cleaning and disin-
fecting cycle to be sure it is: clean, functional, and
undamaged.
D
t
[°C/°F]
[min]
<25/77
3
55/131
10
> 10/50
1
90/194
5
-
-
►
Fig. 13
Water
Chemistry/Note
quality
D–W
-
■
FD–W
Concentrate, alkaline:
– pH ~ 13
– <5 % anionic surfactant
■
working solution 0.5 %
– pH ~ 11*
FD–W
-
FD–W
-
-
mind. 10 min at max. 120 °C
Spray product after every cleaning and disinfection
with oil spray adapter GB600840 (black) approx.
2 sec with Aesculap-STERILIT-Power-Systems oil
spray GB600, see Fig. 13.
en
19