11.1 Operating mode
Operating
Operation with non-periodic load
mode
and speed changes (type S9 pursu-
ant to IEC EN 60034-1)
Drilling (clockwise/counterclock-
wise rotation):
■
60 second application,
60 second pause
■
20 repetitions
■
30 min cooling time
Drilling (oscillation):
■
60 second application,
60 second pause
■
4 repetitions
■
30 min cooling time
Saw mode with GB660R:
■
30 second application,
60 second pause
■
3 repetitions
■
30 min cooling time
11.2 Ambient conditions
Operation
Temperature
10 °C to 27 °C
Without con-
densation
Relative
30 % to 75 %
humidity
Atmospheric
700 hPa to
pressure
1 060 hPa
12. Disposal
Note
The user institution is obliged to process the product
before its disposal, see Validated reprocessing proce-
dure.
►
Storage and
transport
-10 °C to 40 °C
Maximum rec-
ommended
duration of stor-
age is 3 months,
No direct sun-
light
10 % to 90 %
500 hPa to
1 060 hPa
Adhere to national regulations when
disposing of or recycling the product,
its components and its packaging!
The recycling pass can be down-
loaded from the Extranet as a PDF
document under the respective arti-
cle number. (The recycling pass
includes disassembling instructions
for the product, as well as informa-
tion for proper disposal of compo-
nents harmful to the environment.)
Products carrying this symbol are
subject to separate collection of
electrical and electronic devices.
Within the European Union, disposal
is taken care of by the manufacturer
as a free-of-charge service.
Detailed information concerning the disposal of the
product is available through your national
B. Braun/Aesculap agency, see Technical Service.
en
23