Mount cover 11 (sterile) in such a way that it
engages at both cover release buttons 12, see
Fig. 4.
11
Fig. 4
Note
Sterility of the unit is ensured only if the cover is
mounted correctly.
12
Removing the battery
Risk of injury from hot bat-
tery!
Battery may be hot after use in
WARNING
the machine.
►
►
Damage to the battery caused
by knocking it against hard
objects!
CAUTION
►
Damage to, or destruction of
the batteries due to process-
ing!
CAUTION
►
The battery must be removed after the operation and
prior to processing of the unit.
Note
If necessary, the battery removal device can be used to
facilitate removal of the battery, see Fig. 7!
Let the battery cool in the
machine and then remove.
or –
Remove the battery with
the battery removal device
and let it cool.
Only remove battery by
tapping the product on the
palm of the hand.
Do not sterilize batteries.
en