5 – Maintenance
Cleaning, disinfecting, sterilizing
Perform the following maintenance activities on your scanner and accessories regularly.
To ensure maximum hygienic safety for the patient, carefully follow the instructions to prepare the
scanner for use.
To ensure maximum hygienic safety for the patient and to minimize the risk of cross-contamination,
carefully perform the following maintenance activities on your scanner and accessories.
After each patient:
• Clean and disinfect the scanner. See "Clean and Disinfect the Scanner".
• Clean the Scanner Tip, and then perform Autoclave Sterilization (See "Clean the Scanner Tip" and
"Sterilize the Scanner Tip").
There are 2 models of Scanner Tip:
Model
Size
TP102N
Large
TP103N
Small
Clean and disinfect the scanner
General Warnings
WARNING
• Read and follow the warnings and personal protection instructions provided in the Safety Data Sheet
(SDS) for the disinfectant used to process the scanner.
• You must wear gloves while cleaning and disinfecting the scanner.
• The scanner must be disinfected with a recommended intermediate-level disinfectant solution with
tuberculocidal activity between patients.
• DO NOT use a disinfectant containing phenolics or iodophors; doing so will damage the surface
coating of the scanner.
• Never put the scanner in an autoclave device or immerse it in water or the disinfectant solution.
• Excessive fluids can damage the scanner.
• Do not use cotton, cloth, or tissues soaked with disinfectant to disinfect the scanner.
Clean the Scanner
• If the scanner is visibly contaminated with blood and/or body fluids, you must clean it before
disinfecting it.
• To clean the scanner, follow these steps:
– Dampen (do not soak) a lint-free cloth with lukewarm water.
– Remove the blood and/or body fluids with the dampened lint-free cloth.
1 6
NeoScan™ 2000 Installation and operation manual // Maintenance
UDI-DI
(01)06973993441034
(01)06973993441041
Manual
134° C
Cleaning
Sterilization