7.4 Diagnostics/Test menu
The Diagnostics/Test menu shows current and historical information about the sensor. Refer to
Table
4. Push the main menu icon, then select Devices. Select the device and select Device menu
> Diagnostics/Test.
Option
Description
Module
For sensors connected to a pH/ORP module only—Shows the version and the serial
information
number for the sensor module.
Sensor
For sensors connected to a pH/ORP module—Shows the sensor name and the serial
information
number entered by the user.
For sensors connected to an sc digital gateway—Shows the sensor model number, and
the sensor name entered by the user and the sensor serial number. Shows the software
version and driver version installed.
Last calibration
For sensors connected to a pH/ORP module only—Shows the number of days since the
last calibration was done.
Calibration history
For sensors connected to a pH/ORP module—Shows the calibration slope and date of the
previous calibrations.
For sensors connected to an sc digital gateway—Shows the calibration slope and date of
the last calibration.
Reset calibration
For sensors connected to a pH/ORP module only—For service use only
history
Impedance status
For pH sensors only—Refer to
Sensor signals
For pH sensors connected to a pH/ORP module only—Shows the current reading in mV.
(or Signals)
For pH sensors connected to an sc digital gateway—Shows the current reading in mV and
the analog to digital converter counters.
If Impedance status is set to Enabled, shows the active and reference electrode
impedances.
Sensor days
For sensors connected to a pH/ORP module—Shows the number of days that the sensor
(or Counter)
has been in operation.
For sensors connected to an sc digital gateway—Shows the number of days that the
sensor and electrode(s) have been in operation. The Electrode days counter is reset to
zero when the firmware identifies that a defective electrode has been replaced with an
electrode that operates correctly.
To reset the Sensor days counter to zero, select Reset.
Reset the Sensor days counter when the sensor (or salt bridge) is replaced.
7.5 Error list
When an error occurs, the reading on the measurement screen flashes and all outputs are held when
specified in the Controller > Outputs menu. The screen changes to red. The diagnostics bar shows
the error. Push on the diagnostic bar to show the errors and warnings. As an alternative, push the
main menu icon, then select Notifications > Errors.
A list of possible errors is shown in
Error
pH value is too high!
ORP value is too high!
20 English
Table 4 Diagnostics/Test menu
Impedance measurements
Table
5.
Table 5 Error list
Description
The measured pH is > 14.
The measured ORP value is
> 2100 mV.
on page 15.
Resolution
Calibrate or replace the sensor.