6. TROUBLESHOOTING
BEHAVIOR
The Bed Frame does not respond
when Buttons are pressed.
Parts of the Bed Frame remain in
an elevated position after a power
failure.
ASHLEY®
POSSIBLE CAUSE / SOLUTION
1. The Bed Frame is not connected properly. Check the power connection
2. Time-out has been reached due to continuous operation. Do not operate
the Bed Frame for at least 6 minutes. This will reset the time-out.
3. Unplug the Power Converter (L). Wait until the LED goes out, then plug it
back in.
4. Replace the batteries in the Remote Control (AB).
If power cannot be restored, an Emergency Lowering is required. Proceed
as follows:
1. Unplug the Power Converter (L) from the power outlet.
2. Disconnect the Y cord (J) from the DC power cord (G).
3. Connect the Battery Strap (P) to the DC power cord (G).
4. Connect two 9V batteries to the Battery Strap (P).
5. Use the Head DOWN/UP, Foot DOWN/UP or Sleeping Position Buttons to
drive the Bed Frame to its lowest position.
Once Emergency Lowering is complete and power is restored, you can
reconnect the Power Converter. Then, perform a Reference Drive (see Chapter
4) and operate the Bed Frame as normal.