INDICATOR LIGHT ExPLANATION
INDICATOR LIGHT COLOR
None
Standby (green)
Filling (green)
Full (green)
Service (red) and Audible Alert
REMOVING THE IFILL OxYGEN CYLINDER
Do not modify any connections on the personal oxygen station. Never lubricate the connec-
tions. Never use tools of any kind to connect/disconnect the oxygen cylinder and the oxygen
station. Otherwise, severe injury and/or damage may occur.
Use extreme care when handling and filling an oxygen cylinder. Full oxygen cylin ders are
under pressure and can become a projectile if dropped or mishandled.
NOTE– Refer to the General Dangers & Warnings and the Handling Warnings in this man-
ual.
1. Remove the full oxygen cylinder by pushing down on the fill port collar located under the
cylinder nipple connector while using your other hand to steady the cylinder (Figure 4).
2. Lift up on the cylinder to remove from the oxygen fill connector. The green Standby light
will turn on after a short delay.
3. Oxygen will escape from the fill connector for a short time after the cylinder has been
removed. This may be accompanied by a "pop" with a rush of air. This is normal.
4. Perform one (1) of the following steps:
A. If desired, fill another oxygen cylinder. Refer to the DeVilbiss iFill Personal Oxygen
Station Operation Checklist.
B. Place the oxygen fill connector cover onto the fill connector.
NOTE– The oxygen fill connector cover should always be used whenever cylinders are not being
filled.
SETTING THE IFILL OxYGEN CYLINDER ROTARY SELECTOR TO THE
PRESCRIBED SETTING
1. Attach the nasal cannula to the cannula fitting of the iFill oxygen cylinder.
2. Turn the rotary selector to the L/min. setting prescribed by your physician.
Changing the L/min. setting on the rotary selector will affect the dose of oxygen delivered,
DO NOT readjust the L/min. setting unless directed by your physi cian.
3. To turn the flow of oxygen off, turn the rotary selector counterclockwise to the "OFF" posi-
tion.
TROUBLESHOOTING
The following troubleshooting chart will help you analyze and correct minor system malfunc-
tions. If the suggested procedures do not help, call your Sunrise Medical/DeVilbiss homecare
provider. Do not attempt any other maintenance.
To avoid electric shock hazard, do not remove the cabinet. The cabinet should only be
removed by a qualified Sunrise Medical/DeVilbiss homecare technician.
OxYGEN STATION STATUS
Oxygen station is unplugged.
Oxygen station is ready to begin filling cylinders.
Oxygen station is on and cylinder is filling.
Cylinder is full.
Oxygen station is on but cylinder is not filling due to inter-
nal failure.
WARNING
WARNING
WARNING
O P E R AT IO N / T R OU B L E SHO O T I N G
13
H E A D E R
EFFECT
None
Cylinder filling not started
Cylinder is filling
Remove cylinder.
Contact your Sunrise Medical
Provider
Figure 4
SE-535