Descargar Imprimir esta página

DeVilbiss Healthcare iFill 535I Guía De Instrucciones página 14

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 16
E N G L I SH
E N G L I SH
DEVILBISS IFILL PERSONAL OxYGEN STATION
PROBLEM
No indicator lights on.
Red light on.
Excessive fill times OR Green
filling light stays on.
IFILL OxYGEN CYLINDER/REGULATOR
ΠΡΟΒΛΗΜΑ
Hissing sound from iFill oxy-
gen cylinder
iFill oxygen cylinder is on but
oxygen is not being delivered.
TYPICAL QUESTIONS AND ANSWERS
Q.
Why does my iFill make a "pop" sound when I remove the cylinder?
A.
The "pop" sound is pressure being released from the system to make the next cylinder easier to install.
Q.
Why is the coupler only to be cleaned with a clean, dry, lint-free cloth and only when necessary?
A.
Cleaning solutions may be flammable and/or may contaminate the cylinder.
Q.
Why can't I use my old cylinders with the iFill?
A.
The iFill has a proprietary connection as required by the Food and Drug Administration (FDA) and, therefore, other cylinders
are not compatible.
CLEANING/MAINTENANCE
DEVILBISS IFILL PERSONAL OxYGEN STATION FILTER
CAUTION – Do not operate the personal oxygen station without the filters installed.
NOTE– The DeVilbiss iFill Personal Oxygen Station has two (2) filters, one on each side of the cabinet (FIGURE A).
NOTE– Perform this procedure at least ONCE A WEEK.
1. Unplug the personal oxygen station from the wall outlet.
2. Remove the filters.
3. Clean the filters with a vacuum cleaner or wash in warm soapy water and rinse thoroughly.
4. Dry the filters thoroughly before reinstallation.
ExTERIOR CABINET
Electric Shock Hazard: Unplug the DeVilbiss iFill Personal Oxygen Station when cleaning. Do not remove the oxygen station cabi-
net. The cabinet should only be removed by a qualified Sunrise Medical/DeVilbiss technician. Do not apply liquid directly to the
1. Clean the exterior cabinet by using a damp cloth or sponge with a mild house hold cleaner and wipe it dry.
2. Only if necessary, wipe the coupler with a clean, dry, lint-free cloth.
SE-535
T R OU B L E SHO O T I N G / C L E A N I N G / M A I N T E NA N C E
1. Check that the personal oxygen station power cord is properly plugged into the wall outlet.
2. Ensure the fuse in the Mains Power Entry Module is not open.
3. If there are still no indicator lights, contact your Sunrise Medical/DeVilbiss provider.
1. Ensure the iFill oxygen cylinder is connected correctly.
2. Press the start/stop switch. If the Red light remains on, contact your Sunrise Medical/DeVilbiss
provider for service.
3. If the Red light turns on during the filling process, check that the cylinder is connected properly.
Remove the cylinder and reinstall. Ensure the cylinder rotary selector is in the OFF position.
Press the start/stop switch to turn off the Red light and then again to restart the cylinder filling
process. If the Red light turns on again during the filling process, contact your Sunrise Medical/
DeVilbiss provider for service.
1. Ensure the iFill oxygen cylinder rotary selector is set to OFF.
2. Check the fill times shown in Typical Oxygen Cylinder Fill Times chart.
3. Check that the cylinder is connected properly: Press the start/stop switch to return to Standby.
Remove the cylinder and reinstall. Press the start/stop switch to begin filling the cylinder.
4. If problem persists, contact your Sunrise Medical/DeVilbiss provider for service.
1. Turn rotary selector OFF.
2. If hissing sound continues, call your Sunrise Medical/DeVilbiss provider.
1. Refill cylinder.
2. Check that cannula tubing is straight and not pinched. If damaged, replace.
3. Call your Sunrise Medical/DeVilbiss provider for service.
cabinet nor utilize any petroleum- based solvents or cleaning agents.
SOLUTION
ΛΥΣΗ
WARNING
14
H E A D E R

Publicidad

loading