6 - REGENERATION OF THE MASK FOR REUSE ON DIFFERENT
PATIENTS
The mask can be regenerated for up to 20 cycles by one of the following
processes:
• high level thermal disinfection;
• high level chemical disinfection;
• autoclave sterilization at 121°C for a maximum of 30 minutes.
For full detailed mask regeneration instructions, consult the "Guide to
decontamination, disinfection and sterilization" available electronically on the
website medicaldevice.airliquide.com or at your dealer.
PRECAUTIONS
• The mask must always be regenerated before use on a new patient.
• The headgear must be replaced.
7 - TROUBLESHOOTING
PROBLEMS
POSSIBLE CAUSES
The airflow
- Flow generator turned
doesn't reach
off or not working.
the mask.
- Breathing circuit
ventilator not inserted
correctly.
Noisy and
- The mask has not
annoying air
been assembled
leaks.
correctly.
- The mask is not
positioned correctly.
- The size of the mask
may not be correct.
8 - DISPOSAL
The mask components are made of the materials indicated below. They
do not contain hazardous substances and can therefore be disposed of
with ordinary household waste.
ENGLISH
SOLUTIONS
- Switch on or adjust the flow
generator.
- Connect the breathing
circuit ventilator to the flow
generator and the mask.
- Disassemble the mask
and then reassemble it as
described in the user manual.
- Reposition the mask on
your face and adjust the
headgear straps as described
in the user manual.
- Contact your doctor or
therapy technician to find the
right size.
COMPONENT
Cushion (a)
Frame (b)
Blue elbow connector (c)
Headgear (d)
Buckle (e)
9 - TECHNICAL DATA
Regulatory information
Operating pressure range
Operating temperature range
Dead space
Dimensions (without
headgear)
Weight
10 - WARRANTY CERTIFICATE
Air Liquide Medical Systems S.r.l. guarantees its product against material or
manufacturing defects for 180 days from the date of purchase, provided that
the conditions of use in the instructions for use are observed.
If the product is faulty under normal use, Air Liquide Medical Systems S.r.l.
will, at its discretion, repair or replace the faulty product or its components.
The manufacturer shall be responsible for the safety, reliability and
performance of the device if it is used in accordance with the operating
instructions and its intended use.
Keep the original casing for the entire warranty period. The warranty in case
14
MATERIAL
Silicone + Polycarbonate
Polycarbonate
Polycarbonate
Nylon/UBL/Polyurethane
Nylon
CE marking in accordance with
Regulation 2017/745 (EU) and
subsequent implementation amendments.
Class IIa
Main product standard UNI EN ISO
5356-1
4 cmH
O / 32 cmH
O
2
2
+5 °C / +40 °C
Size S: 281 mL
Size M: 293 mL
Size L: 313 mL
Size S: 190 x 110 x 110 mm
Size M: 190 x 110 x 110 mm
Size L: 190 x 110 x 110 mm
Size S: 117 g
Size M: 118 g
Size L: 120 g