Descargar Imprimir esta página

Pari PEP S System Instrucciones De Uso página 57

Ocultar thumbs Ver también para PEP S System:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 162
Chemical cleaning
with disinfection:
EQUIPMENT:
The method has been validated in Europe using:
– Aldehyde-free instrument disinfectant: Bode Bomix
tration: 2%)
Active agent basis: Quaternary ammonium compound
– Application time: 5 minutes
PROCEDURE:
Clean and disinfect the individual parts in a single work step with
a solution prepared according to the manufacturer's instructions.
Info: If the recommended treatment time is exceeded signific-
antly, the plastic parts may take on the smell of the disinfectant.
RINSING:
 CAUTION! Disinfectant residues can cause allergic reac-
tions or irritation of the mucous membrane. Rinse off all parts
thoroughly in running water at about 15 °C for 3 minutes.
Dispose of the used solution (the diluted solution can be dis-
posed of down the drain).
DRYING:
Shake the water out of all of the parts.
Place all parts on a dry, clean and absorbent surface and allow
them to dry completely.
®
 plus (concen-
25

Publicidad

loading

Este manual también es adecuado para:

018018g4200