EN-2
4. OPERATION
Output
With the interface installed, the indicator will operate in one of the following two ways:
Auto Print = OFF, Stable = ON or OFF
Press the Print button will send the display data.
If Stable is set to be ON, the indicator will wait for a stable reading before sending the stable data.
Auto Print = ON, Stable = ON or OFF
The indicator will automatically send data based on the settings in the menu.
If Stable is set to be ON, only stable values will be sent.
Input
The indicator will respond to various commands sent via the interface adapter. Terminate commands with
[CRLF].
Commands
For the full list of commands, please refer to the indicator's instruction manual.
5. CONFIGURATION
The Ethernet menu will not show up until the Ethernet interface is mounted successfully.
Reset
IP Address
Version
Ethernet
Alt Print CMD
Alt Tare CMD
Alt Zero CMD
End
NOTE: / indicates that it is not applicable in the table above.
For printing template setup, please refer to the following table.
Reset
Assignment
Stable Weight Only [Demand]
Print Setup
Content
Layout
End
NOTE: / indicates that it is not applicable in the table above.
MAC Address
The Medium Access Control (MAC) Address of the Ethernet option board is not editable.
Port
The Ethernet option board has a TCP server, with the port of 9761 (unmodifiable).
IP Address
IP is 169.254.1.1 by default.
no, yes
169.254.1.1-9761
1.03
A~Z, P
A~Z, T
A~Z, Z
/
Mode [Auto On Stable]
Time [Interval(seconds)]
no, yes
Demand, Auto On Stable, Auto On Accept, Interval
(seconds), MT-Continuous, OH-Continuous, SICS
Off, On (LFT is forced to be On)
Load, Load and Zero
1~5000
Result, Gross, Net, Tare, Header, Footer, Mode, Unit, Info
Format, Feed
/
Ethernet