RS232 Interface
1. INTRODUCTION
This RS232 interface kit is for use with the Courier 7000 scale.
2. INSTALLATION
1.
Connect the converter to the RS232 interface.
2.
Plug the RS232 interface into the terminal through
the Mini DIN connector.
NOTE: Plug in the RS232 interface when the USB
Type-C is disconnected from the scale.
3. CONFIGURATION
The OHAUS Interface is preset to communicate using the following settings: 9600 baud, 8 bits, no parity, no
handshake. Check the RS232 device parameters required for the printer or computer that is to be connected.
Check the computer or printer documentation if assistance is required. If the parameters do not match, you
must change either the indicator's, the computer's or the printer's settings.
Upon installation of the interface, the indicator will recognize the RS232 interface and add relevant items to the
menu. Configure the indicator to the desired RS232 device and printing parameters; refer to the indicator's user
manual for assistance.
Menu
Sub-Menu
Reset
Baud Rate
Parity
2nd Serial
Stop Bit
2.S.E.r.I.A
(
)
Handshake
Alt Print CMD
Alt Tare CMD
Alt Zero CMD
End
NOTE: Please refer to the indicator's instruction manual for commands and printouts.
Sub-Menu
Options
(In segment)
reset
no, yes
BAud
300, 600, 1200, 2400, 4800,
9600, 19200, 38400, 57600
PAritY
7 Even, 7 Odd, 7 None, 8
None
StOP
1 bit, 2 bit
H.SHAKE
None, Xon/Xoff
ALt.P
'A' ~ 'Z', P
ALt.t
'A' ~ 'Z', T
ALt.Z
'A' ~ 'Z', Z
ENd
/
Converter
Mini DIN connector
Options (in segment)
NO, yes
/
7 EVEN, 7 Odd, 7
NONe, 8 NONE
1 bIt, 2 bIt
NONe, ON-OFF
/
/
/
/
EN-1