4.2 Installation Requirements
This table is translated
Table 1
Soxtec 2050 Automatic Extraction System
Power consumption, Control Unit
Power consumption, Drive Unit
Power supply
Water supply
Weight, Extraction Unit
Weight, Control Unit
Weight, Drive Unit
Dimensions LxDxH, Extraction Unit
Dimensions LxDxH, Control Unit
Dimensions LxDxH, Drive Unit
Ventilation
4.2.1 Input and Output Connections
At the rear:
Control Unit: Mains, DU cable/CU cable, hot plate cable, temperature sensor cable,
air.
Drive Unit: Mains, DU/CU cable, wires.
4.2.2 Fuses
Control Unit:
Internal: 230 V: T 160 mA, 115 V: T 315 mA
Secondary: 230 V: T 500 mA, 115 V: T 500 mA
Drive Unit:
External: 230 V: T 315 mA, 115 V: T 630 mA
4.3 Installation Instructions
The installation should be performed by a service engineer authorised by FOSS An-
alytical AB.
The following text refers only to simpler installation operations and recommenda-
tions:
4.3.1 Fume Hood
Warning
The instrument must only be used in a fume hood which has the main power in-
terlocked to the airflow.
The instrument should be installed in such a way as to comply with the regulations
issued by the local authorities. The Extraction Unit must be installed in a well-venti-
lated fume hood (air flow at least 0.5 m/sec.). The Soxtec 2050 Control Unit and the
Soxtec 2050 Drive Unit are not intended for use in potentially explosive athmosphere
4:2
Soxtec™ 2050 Automatic System
1550 W
40 W
230 V ±10 %, 50-60 Hz, 115V ±10 %, 50-60 Hz
2 l/min tap water at max 20°C
Approx. 30 kg
Approx. 3 kg
Approx. 8 kg
600x380x580 mm
310x240x165 mm
397x280x115 mm
Fume hood. Mains power interlocked to the air-
flow which must be at least 0.5 m/sec.
User Manual 1000 7414 / Rev. 4