3.7.6. Calibration/Test
If a change in calibration of the system is required this should be carried out in the following
manner. Calibration should be carried out using a traceable reference thermometer with an
accuracy of at least 0.1ºC and when stable set CAL value to reference thermometer reading.
Press both buttons for > 3 seconds
Internal probe. Sets Low end of
calibration range.
Internal probe. Sets High end of
calibration range.
External probe. Sets Low end of
calibration range
.
External probe. Sets High end of
calibration range.
LCAL & LCLE max: 20°C
hCAL & hCLE min: 70°C
Restore factory internal calibration
values.
Restore factory external calibration
values
.
Test
Not available to user.