Fault/error code
to
020-272XX-XXX
023-272XX-XXX
024-272XX-XXX
10.3
Ultrasonic height measurement
Fault/error code
symbol appears during
calibration
Ultrasound measuring head
power LED does not light up
Foot silhouettes on the
weighing platform do not illu-
minate
Patient instructions are not
announced
No beeps audible
Measured results are not an-
nounced
178 • seca 555/554/552; seca 550/549; seca 257/256
Cause
One corner of the scale has been
loaded excessively
No GAL value
Cause
Calibration failed
Device is in an undefined state follow-
ing implausible input
Wiring in the ultrasound measuring
head incorrect
Power LED is defective
Device is in an undefined state follow-
ing implausible input
Foot silhouette illumination is defective
Function not activated
Volume set to zero
Loudspeaker is defective
Function not activated
Volume set to zero
Loudspeaker is defective
Function not activated
Remedy
• Distribute weight evenly
è Restarting the device, page 184
•
• If the error recurs, inform seca Service
• Inform seca Service
Inform seca Service
Remedy
• Ensure that no objects or people are in
the immediate vicinity of the device during
calibration
• Ensure that the supplied calibration rod
was used
• Ensure that the calibration rod is posi-
tioned centrally on the foot silhouettes of
the weighing platform
è Restarting the device, page 184
•
• If the error recurs, inform seca Service
Wire the ultrasound measuring head as de-
scribed in the corresponding installation in-
structions
Inform seca Service
è Restarting the device, page 184
•
• If the error recurs, inform seca Service
Inform seca Service
è Activating/deactivating
Activate function
announcement of patient instructions
(Measurement), page 170
è Setting the vol-
Increase volume
ume, page 169
Inform seca Service
è Activating/deactivating
Activate function
beeps, page 170
è Setting the vol-
Increase volume
ume, page 169
Inform seca Service
è Activating/deactivating
Activate function
announcement of measured results (Re-
sults), page 171