Descargar Imprimir esta página

GE ImageQuant LAS 4000 Manual Del Usuario página 81

Publicidad

Mensaje de error
Cannot find the appropriate filter for
the method. Please install an appro-
priate filter.
Menú Option: FlatFrame Calibration
Mensaje de error
The door was opened during exposu-
re. Flat frame calibration failed.
An error occurred during flat frame
calibration for Cy3 tray position 4.
Please check that the correct tray
and calibration plate were used.
Cannot start flat frame calibration
because the CCD calibration for High
Resolution mode does not exist.
Please run CCD calibration first.
Menú Option: CCD Calibration
Mensaje de error
The door was opened during exposu-
re. This image may be inaccurate.
Menú Option: CCD Calibration Pro
Mensaje de error
Cannot use the flat frame calibration
for the following settings: Light none,
Filter No1, FUJINON VRF43LMD lens
Getting Started with ImageQuant LAS 4000 28-9607-59 AB
Significado
El filtro adecuado no está
en su lugar.
Significado
Se abrió la puerta durante
la calibración de flat fra-
me.
Falló la captura de imagen
de la calibración de Flat
Frame.
El CCD no se ha calibrado.
Significado
Se abrió la puerta durante
la exposición; por lo tanto,
es posible que la imagen
no se haya expuesto co-
rrectamente.
Significado
La función FlatFrame Cali-
bration Pro no se puede
usar en el modo quimiolu-
miniscente.
6 Resolución de problemas
6.6 Mensajes de error y contramedidas
Contramedida
Asegúrese de que el filtro co-
rrecto para el método esté en
su lugar y de que esté registra-
do en el software.
Contramedida
Cierre nuevamente la puerta y
vuelva a realizar la calibración
de flat frame.
Verifique que la bandeja de
muestra y la placa de calibra-
ción correctas estén en su lu-
gar y vuelva a calibrar.
Realice una calibración de flat
frame.
Contramedida
Realice nuevamente la calibra-
ción del CCD.
Contramedida
Utilice la opción CCD Calibra-
tion.
81

Publicidad

loading