Mensaje de error
The environmental temperature is
too high to cool the CCD.
Components in ImageQuant LAS
4000 have changed.
Please change the exposure method.
Please change settings in Met-
hod/Tray Position window.
Botón Start: inicio de la exposición
Mensaje de error
The interval time must be greater
that 10 seconds when using the In-
crement/Repetition mode.
The maximum exposure time when
using the EtBr method is 10 minutes.
The maximum exposure time when
using the UV light is 10 minutes.
The maximum interval time is 2
hours.
The maximum exposure time is 30
hours.
The maximum interval time when
using the EtBr method is 60 seconds.
The maximum interval time when
using the UV light is 60 seconds.
Please input the file name using the
program settings window.
Please input a valid exposure time.
Getting Started with ImageQuant LAS 4000 28-9607-59 AB
Significado
La temperatura ambiente
es demasiado alta para
que el CCD se enfríe.
Las configuraciones prede-
terminadas han sido modi-
ficadas.
Significado
El valor de tiempo es inco-
rrecto.
No se proporciona un
nombre de archivo en
Program Settings.
El tiempo de exposición no
es correcto.
6 Resolución de problemas
6.6 Mensajes de error y contramedidas
Contramedida
Apague el ImageQuant LAS
4000, reduzca la temperatura
del aire alrededor del instru-
mento y luego reinicie el Image-
Quant LAS 4000 y el Image-
Quant LAS 4000 Control Softwa-
re.
Cambie el método o la posición
de la bandeja en el diálogo
Method/Tray Position.
Contramedida
Ingrese un valor de tiempo co-
rrecto.
Ingrese un nombre de archivo
válido en el diálogo Program
Settings.
Elija un tiempo de exposición
adecuado.