6. Faults
6. Faults
Personnel: Skilled personnel
Protective equipment: Protective gloves, safety goggles
WARNING!
Physical injuries and damage to property and the environment
caused by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, flammable
or toxic substances) or harmful media (e.g. corrosive, toxic, carcinogenic,
radioactive), there is a danger of physical injuries and damage to
property and the environment.
Should a failure occur, hazardous media with extreme temperatures
(over 55 °C [131 °F]) may be present at the instrument.
▶
For these media, in addition to all standard regulations, the
appropriate existing codes or regulations must also be followed.
▶
Wear the requisite protective equipment, see chapter 2.4 "Personal
protective equipment".
If faults cannot be eliminated by means of the listed measures, the
instrument must be taken out of operation immediately.
▶
Contact the manufacturer.
▶
If a return is needed, please follow the instructions given in chapter
8.1 "Return".
For contact details, see chapter 1 "General information" or the back page
of the operating instructions.
Display
Frozen display
BATTERY LOW
WIKA operating instructions, model CTP1500
Cause
Battery completely exhausted
Display error
Battery completely exhausted
Measures
For inserting new batteries, see
chapter 7.2 "Battery replacement"
For removing and reinserting the
batteries, see chapter 7.2 "Battery
replacement"
If the display remains frozen, send
instrument in for repair
For inserting new batteries, see
chapter 7.2 "Battery replacement"
EN
21