1
Specifications of the of the ViscoPump III
1.1
Notes to the operating manual
The provided operating manual will allow you the proper and safe handling of the product. For maximum
security, observe the safety and warning instructions in the operating manual!
Warning of a general danger:
Non-compliance results (can result) in injury or material damage.
Important information for device use.
Refers to another part of the operating manual.
The menu screens shown in this operating manual serve as an example and may differ from what you
see!
1.2
Intended Use
The ViscoPump III module is a functional unit used to determine absolute and relative viscosity of liquid
media. The module is controlled exclusively through the use of a special program to be run either on a
computer or on a suitable control unit (such as the AVS
forms an integral part a viscosity measuring device and is designed for being built into the respective
device or controller unit.
1.3
Technical Specifications
1.3.1 ViscoPump III
Translation of the legally binding German version
The data given below is only valid if the ViscoPump III module is built into a certified device
®
®
(e.g. AVS
370, AVS
Country of origin:
Measurement
parameters:
Capture of measurement
Measurement
ranges:
Viscosity:
Time:
:
Pumping pressure
Measurement
precision:
Time
measurement:
®
470, AVS
Pro)!
EMC compatibility according to the Council Directive: 2014/30/EU;
applied harmonized standards: EN 61326-1
Low-voltage directive according to the Council Directive 2014/35/EU;
Testing basis EN 61 010-1: for laboratory equipment
RoHS Council Directive 2011/65/EU
FCC Part 15B and ICES 003
Germany, Made in Germany
Flow time in seconds [s]
value:
Flow time: Optoelectronic or thermo-resistive capture of the meniscus passage
through the measurement planes of the viscometers
"pressing" mode 0.35 ... 1800 mm
20 ... 25 °C "sucking" mode 0.35 ... 5800 mm
approx.
20 ... 25 °C
5 s to 9999.99 s, Resolution
Fully automatically controlled "sucking" action to approx. -160 mbar
Fully automatically controlled "pressing" action to approx. +160 mbar
Precision (reproducibility and comparability) in accordance with DIN 51562, Part 1
± 0.01 s ± 1 digit, but not more accurate than 0.01 %
measurement incertainty in the determination of absolute, kinematic viscosity
furthermore depends on the incertainty of the numerical value of the viscometer
constants and the measurement conditions, especially as concerns the
measurement temperature.
®
470 Viscosity Measuring Unit). The module
/s (cSt) measuring temperature of approx.
2
/s (cSt) measuring temperature of
2
0.01 s
27
(Release: 18. Oct 2021)