62
1.3
Technical Specifications
Piston burette TITRONIC
1.3.1
Translation of the legally binding German version
EMC compatibility according to the Council Directive: 2014/30/EU;
applied harmonized standards: EN 61326-1
Low-voltage directive according to the Council Directive 2014/35/EU;
Testing basis EN 61 010-1: for laboratory equipment
RoHS Council Directive 2011/65/EU
FCC Part 15B and ICES 003
Country of origin: Germany, Made in Germany
The following solvents/titration reagents are allowed to be used:
 All common titration solutions.
 As reagent water and all non-aggressive non-organic and organic fluids are allowed.
 If using combustible fluids fire please adhere to the Guidelines for Explosion Protection and Prevention of the
chemical industry.
 For fluids with higher viscosity ( 5 mm
speed can be adjusted.
 Fluids with viscosity over 20mm
Display:
3.5 inches -1/4 VGA TFT display with 320x240 pixels
by external multi-range power supply from 100 – 240 V, 50/60 Hz
Power supply:
Input voltage: 12 Volt DC, 2500 mA
Power consumption 30 W
Corresponds to protection class III:
Protection class for dust and humidity IP 50 according to DIN 40 050
manufacturer!
RS-232-C Interface:
Daisy Chain function available
Data bits:
Stop bit:
Start bit:
Parity:
Baud rate:
Address:
RS-232-1
for computer, input Daisy Chain
RS-232-2
devices of SI Analytics
- Titrator TitroLine
- Sample Changer TW alpha plus, TW 7400
- Piston burette TITRONIC
- Balances of the types Mettler, Sartorius, Kern, Ohaus, (for more, please contact us)
- Exit Daisy-Chain
USB Interface:
1 x USB-type-A and 1 x USB-type-B
USB-type A
for connecting of USB keyboard, - printer, - manual controller, - data media
(e.g. USB stick) and USB-Hub
USB-type B
for connecting a PC
®
300
2
/s), lower boiling point or affinity to outgas, the filling and dosage
2
/s cannot be dosed.
Only use the power supply TZ 1853 or a power supply approved by the
adjustable, 7 or 8 Bit (default: 8 Bit)
adjustable, 1 or 2 Bit (default: 1 Bit)
static 1 Bit
adjustable: even / odd / none
adjustable: 1200, 2400, 4800, 9600, 19200 (Default 4800 baud)
adjustable, (0 to 15, default: 01)
®
:
®
7000 / 7500 / 7500 KF / 7750 / 7800
®
300 and 500, TITRONIC
(Release: 07. July 2020)
®
®
110 plus, TITRONIC
universal,