4.5.2.2 Sample weight and volume (sample quantity)
The Sample Quantity (W) item (Fig. 52) is used to select whether one is wishing to use a sample weight or a
sample volume for titration or solution preparation (Fig. 53).
Fig. 52
Fig. 53
You have the following options:
«Manual sample weight»: The sample weight is enquired by a prompt at the start of the method and
manually input.
«Automatic sample weight»: The sample weight is automatically transferred by a connected balance.
«Fixed sample weight»: A fixed sample weight is input in g. This weight will then automatically be used
for each start of the method.
«Manual sample volume»: The sample volume in ml is prompted at the start of the method and
manually input.
«Fixed sample volume»: A fixed sample volume is input in ml. This volume will then automatically be
used for each test of the method.