A Open flow valve on the O
flowmeter or O
2
giFlow and set the oxygen flow necessary for the
patient on the device.
B Only O
flowmeter: Read flow on upper rim of
2
ball.
The nebulizer produces a fine mist.
Shut down
Close flow valve on the O
flowmeter or O
2
giFlow.
Remove inhalation device or insufflation device.
12
Reprocessing
Di-
2
Stripping down
Di-
2
A Disconnect humidifier bottle and empty.
Disconnect O
flowmeter or O
2
housing.
B Remove nebulizer.
C Remove support ring from humidifier bottle.
D Disconnect jet
Cleaning and Disinfection
Observe the hygiene regulations of the hospital.
Clean O
device and insufflation device following the rele-
vant Instructions for Use.
CAUTION
Risk of stress cracking. Do not use disinfectants or
D
cleaning agents containing alcohol.
Reprocessing Procedure
B
Testing of procedures and agents
The cleaning and disinfection of medical products
C
has been tested with the following procedures and
agents. The following agents showed good material
compatibility at the time of the test:
Manual cleaning: Neodisher LM2 by Dr. Weigert
Manual disinfection: Korsolex extra by Bode
A
Machine cleaning: Neodisher MediClean by
Dr. Weigert
Machine disinfection: thermal disinfection at
93 °C (199.4 °F), 10 minutes
Sterilization: hot steam sterilization at 134 °C
(273.2 °F), 5 minutes, fractional vacuum
DigiFlow from
Manual cleaning
2
Manual cleaning should preferrably be carried out
under running water or with commercially available
cleaning agents based on mild alkaline compounds.
Instructions for Use Humidifier Nebulizer
flowmeter or O
DigiFlow, inhalation
2
2