Warning-Score result
A subscore is given for each parameter based on the measured or entered value. When all the required
parameters are entered or measured, the subscores are added together to calculate the total early
warning score. The total score has a color coding to indicate associated severity level. Select Criteria to
check the score criteria.
Select Menu > Score > Setting > Common and set Display Result on the Main Interface to On or Off.
If It is set to On, the warning score result will be displayed in the main interface. Click Details to show the
detailed score result.
Select Menu > Score > Setting > Alarm and set the upper limit of the specific score type. When Switch
is set to On and the total score is beyond the threshold, prompt is given.
EWS
NEWS
NEWS2
MEWS
PEWS
Warning-Score trend table
Trend table provides the monitored patient's scores during a period of time; it includes score time, score
parameters and score. To check the trend table, click Trend Table button in Warning-Score interface. At
least 1200 groups of trend can be reviewed for each warning score system.
If User Maintain > Network > Real-time Upload is turned on, the score data will be uploaded in real time.
Otherwise, click Upload in trend table to manually upload the selected score data.
Delete Warning-Score criteria Chapter
Using Battery
Battery safety information
WARNING
Before using the rechargeable lithium-ion battery (hereinafter called lithium battery), be sure to read the
user manual and safety precautions thoroughly.
WARNING
The lithium battery can only be used for this device.
12
Adjustable Range of Upper Limit
1-17
1-20
1-21
1-15
1-11
Default
4
5
5
4
3
Vista 120 SC – Supplement