11.1 Basic data
"m.v." means a standard measured value.
True RMS measurement for AC current
Range
Resolution
6.000 mA
60.00 mA
600.0 mA
6.000 A
60.00 A
100.0 A
•
All AC voltage ranges are specified from 5% to 100% of range
•
Frequency range: 50 Hz...1000 Hz
•
Overload protection: 120 A
AC current measurement (True RMS) the low-pass filter
Range
6.000 mA
60.00 mA
600.0 mA
6.000 A
60.00 A
100.0 A
•
All AC current ranges are specified from 5% to 100% of range
•
Frequency range: 50 Hz...60 Hz
•
Overload protection: 120 A
40
11 Technical data
for f = 50 Hz...60 Hz
0.001 mA
(1.5% m.v. + 8 digits)
0.01 mA
0.1 mA
(1.0% m.v. + 5 digits)
0.001 A
0.01 A
(1.0% m.v. + 8 digits)
0.1 A
CMP-100 – USER MANUAL
Accuracy
(all waveforms)
Resolution
0.001 mA
0.01 mA
0.1 mA
0.001 A
0.01 A
0.1 A
Accuracy
for f = 60 Hz...1 kHz
(sine waveforms)
(3.5% m.v. + 8 digits)
(3.0% m.v. + 8 digits)
(3.0% m.v. + 5 digits)
(3.0% m.v. + 8 digits)
Accuracy
for f = 50...60 Hz
(2.0% m.v. + 8 digits)
(1.5% m.v. + 8 digits)