Operating Instructions Midwest Tradition
TB/Pro TBF / Midwest Tradition
TL/Pro TLF
7 Conditioning
3. Blow the high-speed handpiece out with max. 3 bar.
4. Maintain mechanical parts manually [ → 41].
5. Maintain the push button chuck manually [ → 43].
6. Pack the high-speed handpiece in packaging material suitable for
sterilization and storage. e.g. paper/plastic composite packaging.
7. Conduct sterilization [ → 47].
7.3
Manual cleaning
IMPORTANT: Manual conditioning is possible in exceptional cases if
the national/regional regulations are followed. The national/regional
regulations are to be checked before.
NOTICE! Condition immediately, or at the latest, one hour after
treatment.
NOTICE! Never clean in an ultrasound bath!
NOTICE! Never immerse in disinfectant solution!
39
66 27 587 D3673.201.02.01.02