Non-contact forehead thermometer FT 90
EN
NOTICE:
• Do not use mobile telephones, microwaves, and other devices which generate
strong electrical or electromagnetic fields near the unit. This may result in incorrect
operation of the unit.
• Do not subject the device to mechanical impacts.
• Portable and mobile HF communication systems may interfere with this device.
For more details, please contact our customer service at the address indicated
or see the enclosed information sheet.
• Do not press the buttons with excessive force or with any pointed objects.
• Never attempt to repair the instrument or adjust it yourself. This may damage the
unit and impair the functions. If you need to have the unit repaired, please contact
our customer service. Please see warranty for service contact. Before submitting any
complaint, first check the batteries and replace them if necessary.
• Changes or modifications to the device will nullify the user warranty.
Briefl y press the On/Off button
After a brief self-test and two short beeps the device is ready for forehead temperature
measurement. The device is in forehead temperature mode. This is indicated by the
symbol.
Beurer tip
Remember that
• physical activity, increased perspiration on the forehead, taking vasoconstrictive
medication and skin irritations can distort the result.
• the forehead must be free from perspiration and cosmetics.
• before taking a reading, please remain indoors at room temperature for 5 min-
utes avoiding exercise or taking a bath for 30 min prior reading.
• in case of doubt, please measure several times and take the highest measure-
ment value.
14
.