Standard methods KF
Titer 1-Component (liquid standard)
Titer 1-Component (solid standard)
Titer 1-Component (water)
Titer 2-Component (liquid standard)
Titer 2-Component (solid standard)
Titer 2-Component (water)
Sample 1-Component
Sample 2-Component
Statistics are switched on. The mean value of the titer in mg/ml is automatically saved in the attachment.
It is then used automatically in the sample titration.
The results of the sample titration are calculated in %.
If needed, the unit can be converted into other units of measure, such as ppm.
Application
Determination of the concentration of the titration agent. Suitable
for 1-component reagents.Standard is a liquid standard in
ampoules with a concentration of 10 mg/g.
Determination of the concentration of the titration agent. Suitable
for 1-component reagents Standard is the standard substance
sodium tartrat dihydrate with a water amount of 15.66 %.
Determination of the concentration of the titration agent. Suitable
for 1-component reagents Standard is pure water
Determination of the concentration of the titration agent. Suitable
for 2-component reagents. Standard is a liquid standard in
ampoules with a concentration of 10 mg/g.
Determination of the concentration of the titration agent. Suitable
for 2-component reagents Standard is the standard substance
sodium tartrat dihydrate with a water amount of 15.66 %.
Determination of the concentration of the titration agent. Suitable
for 1-component reagents Standard is pure water
Method for sample titrations with 1-component reagents
Method for sample titrations with 2-component reagents