Backrest Function (Br) - wissner-bosserhoff Carisma Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 302
Caution! All electrical functions could have faults - unless you lock them
and just release them for the duration of the function.
Caution! Other devices that emit electromagnetic radiation (e.g. radio tele-
in general it applies that bed functions not required have to be locked for
reasons of safety. The use of such devices in the vicinity should be avoided
as far as possible.
Caution! When moving the bed to different positions, it must be ensured
that no persons, body parts or objects are in the adjustment area or bet-
Caution! Since the system can be run via battery operation (optional), it is
not enough just to pull out the plug in case of the motors functioning incor-
rectly. If an incorrect function occurs, please switch off functions via the
control box and contact our Customer Service (battery operation option).
8.1.

BACKREST FUNCTION (BR)

The BR moves towards the head end of the bed during the upward motion. This func-
becoming restricted in the upper body area.
The natural movement of the body is thus supported by this technology.
Caution! Do not operate under load. When operating the mechanical emer-
mattress restricting bar or lifting pole to avoid a sudden lowering.
Recommendation: Handling can be simplified by having the patient/resident firmly hold
the lifting pole trapeze (if available) so that there is no load at all on the backrest.
Handling: Hold the BR in the set position and remove the load from it slightly. Turn the
red lever (on the side of the motor unit covering) in the direction of the backrest.
Handling: Hold the thigh rest in the set position and remove the load from it slightly. Turn
the red lever (on the side of the motor unit covering) in the direction of the thigh rest.
BR/TR (Emergency Release)
Direction of rotation
Page 71