Scan Measurement

Virtual aperture technology
During a scan measurement the i1Pro device is performing 200 measurements per
second. The automatic patch detection of the device identifies useable measurements
made on a patch and unusable measurements made between two patches. Valid
measurements on a patch are averaged and the device reports the averaged result to
the software. Thanks to this technology the virtual aperture of the i1Pro device adapts to
the length of a patch. For best measurement results the length of the patches on your
test chart should be selected based on the resolution of your printer. For printers with
lower resolution or a grainy screening you should increase the length of the patches on
your test chart.
16
Measurement conditions
The i1Pro measurement device is equipped with a fixed UVcut filter and supports only a
single measurement condition.
Single Pass Measurement is required for measurement condition
ISO 13655 M2 | Illumination with UV excluded (UVcut)
Status Indicators on the i1Pro device guides you through the measurement
The
process.
EN