Operation
User navigation
Refer to the controller documentation for keypad description and
navigation information.
Configure the sensor
Use the Configure menu to enter identification information for the sensor
and to change options for data handling and storage.
1. Push the MENU key and select Sensor Setup, [Select Sensor],
Configure.
2. Use the arrow keys to select an option and push ENTER. To enter
numbers, characters or punctuation, push and hold the UP or DOWN
arrow keys. Push the RIGHT arrow key to advance to the next space.
Option
Description
SENSOR NAME
Changes the name that corresponds to the sensor on
the top of the measure screen. The name is limited to
10 characters in any combination of letters, numbers,
spaces or punctuation.
SERIAL NUMBER
Sets the serial number of the sensor, limited to 16
characters in any combination of letters, numbers,
spaces or punctuation.
FLOW FORMAT
Sets the number of decimal places that are shown on
the measure screen to XXXX, XXX.X or XX.XX.
MEASUREMENT
Sets the units for the selected measurement—mm,
UNITS
m, in. or ft
TEMPERATURE
Sets the temperature units to °C (default) or °F.
UNITS
VOLUME UNITS
Sets the units for volume (e.g., liters).
TOTALIZER MODE
Sets the totalizer volume to Auto (default) or Manual.
TOTALIZER RESET
Resets the totalizer volume. Appears only if the
totalizer mode is set to Manual.
Option
Description
GAUGE
Sets the type of gauge and edits values for the gauge
PARAMETERS
type. For more information, see
parameters
SET SYSTEM
Configures the system to define specific sensor
application data.
SET FILTER
Sets a time constant to increase signal stability. The
time constant calculates the average value during a
specified time—0 (no effect, default) to 60 seconds
(average of signal value for 60 seconds). The filter
increases the time for the sensor signal to respond to
actual changes in the process.
FLOW LOG
Sets the time interval for data storage in the data log
INTERVAL
—5, 30 seconds, 1, 2, 5, 10, 15 (default), 30, 60
minutes.
VOLUME LOG
Sets the time interval for data storage in the data log
INTERVAL
—5, 30 seconds, 1, 2, 5, 10, 15 (default), 30, 60
minutes.
RESET DEFAULTS
Sets the configuration menu to the default settings.
All sensor information is lost.
Configure gauge parameters
Configure the controller for the type of open channel flow gauge structure
that is used. The controller has built-in dimension table algorithms for most
common flumes and weirs. If the structure is not covered by the built-in
tables, create a user-definable flow/depth curve (between 3 and 30 points)
to configure the structure.
1. Push the MENU key and select Sensor Setup, [Select Sensor],
Configure.
2. Select GAUGE PARAMETERS and push ENTER.
3. Select GAUGE TYPE and push ENTER.
Configure gauge
on page 9.
English 9