Connect The Sensor To The Module - Hach LANGE Manual Del Usuario

Ocultar thumbs Ver también para LANGE:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 62
Figure 3 Additional mounting examples
1 Top mount option
2 Side mount option 1

Connect the sensor to the module

D A N G E R
Electrocution Hazard. High voltage wiring for the controller is conducted behind the
high voltage barrier in the controller enclosure. The barrier must remain in place
except when installing modules, or when a qualified installation technician is wiring
for power, relays or analog and network cards.
6 English
3 Side mount option 2
4 Top view of side mount option 2
W A R N I N G
Potential Electrocution Hazard. Always disconnect power to the
instrument when making electrical connections.
W A R N I N G
Potential Electrocution Hazard. Only qualified personnel should conduct
the tasks described in this section of the manual.
N O T I C E
Potential Instrument Damage. Delicate internal electronic components
can be damaged by static electricity, resulting in degraded performance
or eventual failure.
An ultrasonic sensor can be connected to the flow module. To install the
module and connect the sensor, refer to the illustrated steps and
Note: The module cannot be connected to multiple sensor types at the same time.
Table 3 Ultrasonic wiring table
Connector
Sensor
6-pin
Ultrasonic
Sensor shield wires – Connect all sensor ground/shield
wires to the sc200 enclosure grounding screws.
Table
3.
Connector
Wire color
pin
1
2
3
Red
4
Green
5
Yellow
6
Blue
Black

Publicidad

Tabla de contenido
loading

Tabla de contenido