If the heartbeat is irregular, it may take
up to 30 seconds to find 6 representative
pulse waveforms. In extreme situations,
it may not be possible to obtain a result
and the test should be repeated. If this
still
does
not
work,
increasing
the
recording time in the settings should
solve the problem and confirming that
the minimum number of pulses is set to
6, page 12.
Once a minimum of 6 waveforms have
been collected and averaged, the result is
displayed.
Select 'Accept' or 'Reject' to proceed to
the Spot Check Menu or select 'Enter BP'
to enter the subjects BP.
Select 'Comments' to enter a comment
from a user defined 'pick' list for this
specific measurement.
Comments are
pre-configured
in
the
Spot-Check
customisation section, page 12.