Avoid operating the PulseTrace PCA2 in dusty conditions or
•
near heating appliances or radiators.
Do not direct the finger transducer towards a strong light
•
source whilst operating the PulseTrace PCA2.
Check the AC charger for compatibility with local power
•
rating.
Cleaning the Transducer
The Photo-Plethysmography transducer requires no routine
maintenance or servicing.
The transducer and cable (disconnected from the interface box) may
be cleaned using a soft cloth moistened with a solution such as 70%
isopropyl alcohol. Secondly, using another clean cloth saturated with
distilled water, wipe clean all surfaces. Carefully avoid damaging the
surface of the finger pads. Allow the finger probe and cable to dry
thoroughly before use.
Caution:
Do not irradiate, autoclave, soak or immerse in any
solutions.
Caution:
Do not use undiluted bleach.
If a low level disinfection is required, use a 1:10 bleach
solution (5%-5.25% sodium hypochlorite).
Caution:
Do not expose connector pins to cleaning solution as
this may damage sensor.
Servicing
There is no routine maintenance required for the PulseTrace PCA2
and there are no user serviceable parts in this device. Please return
the unit to CareFusion or an authorised agent if servicing is required.