Sensor settings

The following parameters can be changed on the sensors:
Name
Alarm threshold A1 (in measuring unit)
Alarm threshold A2 (in measuring unit)
1)
Interpretation type
Alarm threshold STEL
1)
(in measuring unit)
Alarm threshold TWA
1)
(in measuring unit)
1) Interpretation only if the sensor is designed for this.

Check of parameters

To ensure that the values were correctly transferred to the gas detection instrument:
Press the data from X-am 1/2/5x00 in Dräger CC Vision.
Check parameters.
18
Range
0 - A2
A1 – measuring range end value
Inactive, TWA, STEL, TWA+STEL
0 – measuring range end value
0 – measuring range end value
Operation

Preparations for operation

Before using the instrument for the first time, insert a charged NiMH T4
power pack or batteries approved by Dräger, see chapter "Changing the
batteries" on page 22.
The instrument is now ready for operation.
To reduce the risk of ignition of a flammable or explosive atmosphere, strictly
adhere to the following warning statements:
Only use power pack types ABT 01xx, HBT 00xx or HBT 01xx. See the marking
on the rechargeable battery for permitted rechargeable batteries and the
corresponding temperature class.
Substitution of components may impair intrinsic safety.

Switching on the instrument

 Hold down the [OK] key for approx. 3 seconds until the countdown » 3 . 2 . 1 «
shown on the display has elapsed.
– All the display segments, including the visual, audible and vibration alarms,
are activated for a short time.
– The software version is displayed.
– The instrument performs a self-test.
– The sensor that is up next for adjustment is displayed with the remaining days
until the next adjustment, e.g. » Ex %LEL CAL 20 «.
– The time until the bump test interval elapses is displayed in days, e.g. » bt 123 «.
– All alarm thresholds A1 and A2 as well as »
for all toxic gases (e.g. H
S or CO) are displayed consecutively.
2
1) Only when activated in the instrument configuration. Delivery condition: not activated.
WARNING
1)
« (TWA)
and »
« (STEL)
1)