Menus and settings

6 Menus and settings

6.1 Menu structure of pH/ion

1.
2.
6.2 Menu structure of DO
1.
2.
3.

6.3 Temperature settings

1. Set MTC temperature
If the meter does not detect a temperature probe, MTC appears on the display. In this case the sam-
ple temperature should be entered manually. An MTC value between -30 °C and 130 °C can be en-
tered.
2. Temperature unit
Select the temperature unit: °C or °F. The temperature value is automatically converted between the
two units.
6.4 pH/ion calibration settings
1. Buffer groups/standards
1. Predefined pH buffer groups
One of seven predefined buffer groups can be selected:
B1
B2
B3
B4
B5
B6
B7
Temperature tables for these buffers are programmed in the meter and can be found in the "Appendix".
2. Customized pH buffer group
A set of user­defined pH buffers with up to 5 different temperatures for each buffer can be created. The
tem per a ture difference between pH buffers must be at least 5 °C and the difference between the pH values
must be at least 1.
When switching from predefined buffer group to customized buffer group, press Save in the table even if
no values have changed.
22
Temperature Settings
1. Set MTC Temperature
2. Temperature Unit
Calibration Settings
1. Buffer Group/Standards
2. Calibration Mode
3. Calibration Reminder
Continue at top of table
Temperature Settings
1. Set MTC Temperature
2. Temperature Unit
Calibration Reminder
Measurement Settings
1. Salinity Compensation
Continue at top of table
1.68
4.01
7.00
2.00
4.01
9.00
2.00
4.00
7.00
1.679
4.008
6.865
1.680
4.008
6.865
1.09
4.65
6.79
1.680
4.003
6.864
3.
4.
5.
6.
4.
5.
6.
10.01
(at 25°C)
9.21
11.00
(at 25°C)
9.00
12.00
(at 20°C)
9.180
(at 25°C)
9.184
12.454
(at 25°C)
9.23
12.75
(at 25°C)
9.182
12.460
(at 25°C)
Measurement Settings
1. Measurement Resolution
2. Stability Criterion
3. Ion Measurement Unit
4. Rel. mV Offset
Endpoint Formats
Measurement Limits
Sensor ID/SN
2. Barometric Compensation
3. Barometric Unit
Endpoint Formats
Measurement Limits
Sensor ID/SN
Mettler US
Mettler Europe
Standard Merck buffer
JIS Z 8802
DIN19266
DIN19267
Chinese
pH/ORP/Ion/DO meter SG68