Calibration Data - Mettler Toledo SevenGo Duo pro SG68 Instrucciones De Manejo

Ph/orp/ion/do
Ocultar thumbs Ver también para SevenGo Duo pro SG68:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Data management
2. Transfer
All or partially stored measurement data can be transferred by filtering the measurement data. The filter
works as described above in "1. Review".
Press Trans to send the filtered measurement data over the IR interface to the printer or PC.
3. Delete
All or partially stored measurement data can be deleted by filtering the measurement data. The filter works
as described above in "1. Review".
Note
Deletion is protected by a PIN. Upon delivery, the PIN is set to 000000. Change the PIN code to prevent
unau tho rized access.

7.3 Calibration data

Calibration data can be reviewed, transferred and deleted. Up to 5 calibrations per sensor ID are stored in
the memory.
Review
a) Select between the sensor types: pH, ion or DO.
b) Press Review.
c) Select a sensor ID from the list and press Review.
d) Press
— or —
Press and hold CAL for 3 seconds in the single channel measurement screen.
e) Press Trans to send the displayed calibration data over the IR interface to a printer or PC.
Transfer
a) Select between the sensor types: pH, ion or DO.
b) Press Trans.
c) Select a sensor ID from the list and press Trans.
d) Press
e) Press Trans to send all the calibration data of the selected sensor ID over the IR interface to a print-
er or PC.
Delete
a) Select between the sensor types: pH, ion or DO.
b) Press Delete.
c) Select a sensor ID from the list and press Delete.
d) Press Yes when the message "This will delete all your selected data. Please confirm." appears
— or —
Press No to cancel and exit.
Notes
An active sensor ID cannot be deleted.
This menu is protected by a deletion PIN code. Upon delivery, the PIN code is set to 000000.
Change the PIN code to prevent unauthorized access.
28
A list of calibrated sensor IDs appears.
and
to navigate between the previous or next calibration data sets.
The current calibration data is displayed.
A list of calibrated sensor IDs appears.
and
to navigate between the previous or next calibration data sets.
A list of sensor IDs appears.
After deletion, the sensor ID disappears from the list in the sensor ID menu.
pH/ORP/Ion/DO meter SG68

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Sevengo duo proSg68

Tabla de contenido