Maintenance; Technical Data - IKA MS 3 control Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Maintenance

The appliance is maintenance-free.
For cleaning, disconnect the main plug!
Only use cleansing agents which have been recommended by IKA. Use to
remove:
Dyes
isopropyl alcohol
Construction materials
water containing tenside/ isopropyl alcohol
Cosmetics
water containing tenside/ isopropyl alcohol
Foodstuffs
water containing tenside
Fuels
water containing tenside
- Do not allow moisture to get into the appliance when cleaning
- Wear protective gloves during cleaning the devices.
- Before using another than the recommended method for cleaning or
decontamination, the user must ascertain with IKA that this method does
not destroy the instrument.
Spare parts order
When ordering spare parts, please give:
- Machine type
- Manufacturing number, see type plate
- Item and designation of the spare part, see spare parts diagram and spare
parts list, see www.ika.com.
Repair
Please only send devices in for repair that have been cleaned and are
free of materials which might present health hazards.
For this, use the "certificate of compliance" form which you can obtain
from IKA or can download a version for printing from the IKA website at
www.ika.com.
If your appliance requires repair, return it in its original packaging. Storage
packaging is not sufficient when sending the device - also use appropriate
transport packaging.
12

Technical data

Power supply
Input
Output
Protection class
Shaker
Operating voltage
Power consumption, normal operation W
Power consumption, standby operation W
Motor-output power
Drive
Speed range
Speed adjustment
Speed display
Agitation stroke
Shaking motion
Perm duration of operation
Timer mode seconds
mode minutes
Timer display
Perm ambient temperature
Perm. relative humidity
Protection type acc. to DIN EN 60529
Contamination level
Overvoltage category
Operation at a terrestrial altitude
Interface
V
100 - 240
A
0,8
Hz
50-60
Vdc
24
W
24 (Limited power source)
2 (double insulated)
Vdc
24
mA
800
20
2
W
8
EC - motor
rpm
0/100 - 3000
adjustable in tens steps
rotating knob in front
of the device
digital
mm
4,5
horizontal, circular
%
100
1 sec ... 59 min 59 sec
1 min ... 59 h 59 min
digital
°C
+5 - +40
%
80
IP 21
2
II
m
max. 2000
RS 232